Chapter 5 (Art. 51 – 60) Classification and conformity assessment

Article 52 – Conformity assessment procedures

Prior to placing a device on the market, manufacturers shall undertake an assessment of the conformity of that device, in accordance with the applicable conformity assessment procedures set out in to
.
Prior to putting into service a device that is not placed on the market, manufacturers shall undertake an assessment of the conformity of that device, in accordance with the applicable conformity assessment procedures set out in to
.
Manufacturers of class III devices, other than custom-made or investigational devices, shall be subject to a conformity assessment as specified in . Alternatively, the manufacturer may choose to apply a conformity assessment as specified in coupled with a conformity assessment as specified in .
Manufacturers of class IIb devices, other than custom-made or investigational devices, shall be subject to a conformity assessment as specified in Chapters I and III of , and including an assessment of the technical documentation as specified in Section 4 of that Annex of at least one representative device per generic device group.
However, for class IIb implantable devices, except sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors, the assessment of the technical documentation as specified in Section 4 of shall apply for every device.
Alternatively, the manufacturer may choose to apply a conformity assessment based on type examination as specified in coupled with a conformity assessment based on product conformity verification as specified in .
Where justified in view of well-established technologies, similar to those used in the exempted devices listed in the second subparagraph of paragraph 4 of this Article, being used in other class IIb implantable devices, or where justified in order to protect the health and safety of patients, users or other persons or other aspects of public health, the Commission is empowered to adopt delegated acts in accordance with to amend that list by adding other types of class IIb implantable devices to that list or removing devices therefrom.
Manufacturers of class IIa devices, other than custom-made or investigational devices, shall be subject to a conformity assessment as specified in Chapters I and III of , and including an assessment of the technical documentation as specified in Section 4 of that Annex of at least one representative device for each category of devices.
Alternatively, the manufacturer may choose to draw up the technical documentation set out in and
coupled with a conformity assessment as specified in Section 10 or Section 18 of . The assessment of the technical documentation shall apply for at least one representative device for each category of devices.
Manufacturers of class I devices, other than custom-made or investigational devices, shall declare the conformity of their products by issuing the EU declaration of conformity referred to in after drawing up the technical documentation set out in and
. If those devices are placed on the market in sterile condition, have a measuring function or are reusable surgical instruments, the manufacturer shall apply the procedures set out in Chapters I and III of , or in Part A of . However, the involvement of the notified body in those procedures shall be limited:
in the case of devices placed on the market in sterile condition, to the aspects relating to establishing, securing and maintaining sterile conditions;
in the case of devices with a measuring function, to the aspects relating to the conformity of the devices with the metrological requirements;
in the case of reusable surgical instruments, to the aspects relating to the reuse of the device, in particular cleaning, disinfection, sterilization, maintenance and functional testing and the related instructions for use.
Manufacturers of custom-made devices shall follow the procedure set out in and draw up the statement set out in Section 1 of that Annex before placing such devices on the market.
In addition to the procedure applicable pursuant to the first subparagraph, manufacturers of class III custom-made implantable devices shall be subject to the conformity assessment as specified in Chapter I of . Alternatively, the manufacturer may choose to apply a conformity assessment as specified in Part A of .
In addition to the procedures applicable pursuant to paragraph 3, 4, 6, or 7 of this Article, in the case of devices referred to in the first subparagraph of , the procedure specified in Section 5.2 of or Section 6 of , as applicable, shall also apply.
In addition to the procedures applicable pursuant to paragraph 3, 4, 6, or 7 of this Article, in the case of devices that are covered by this Regulation in accordance with point (f) or (g) of and with the first subparagraph of , the procedure specified in Section 5.3 of or Section 6 of , as applicable, shall also apply.
In addition to the procedures applicable pursuant to paragraph 3, 4, 6, or 7, in the case of devices that are composed of substances or of combinations of substances that are intended to be introduced into the human body via a body orifice or applied to the skin and that are absorbed by or locally dispersed in the human body, the procedure specified in Section 5.4 of or Section 6 of , as applicable, shall also apply.
The Member State in which the notified body is established may require that all or certain documents, including the technical documentation, audit, assessment and inspection reports, relating to the procedures referred to in paragraphs 1 to 7 and 9 to 11 be made available in an official Union language(s) determined by that Member State. In the absence of such requirement, those documents shall be available in any official Union language acceptable to the notified body.
Investigational devices shall be subject to the requirements set out in to
.
The Commission may, by means of implementing acts, specify detailed arrangements and procedural aspects with a view to ensuring the harmonised application of the conformity assessment procedures by the notified bodies for any of the following aspects:
the frequency and the sampling basis of the assessment of the technical documentation on a representative basis as set out in the third paragraph of Section 2.3 and in Section 3.5 of in the case of class IIa and class IIb devices, and in Section 10.2 of in the case of class IIa devices;
the minimum frequency of unannounced on-site audits and sample tests to be conducted by notified bodies in accordance with Section 3.4 of , taking into account the risk-class and the type of device;
the physical, laboratory or other tests to be carried out by notified bodies in the context of sample tests, assessment of the technical documentation and type examination in accordance with Sections 3.4 and 4.3 of , Section 3 of and Section 15 of .
The implementing acts referred to in the first subparagraph shall be adopted in accordance with the examination procedure referred to in .
Want to print your doc?
This is not the way.
Try clicking the ⋯ next to your doc name or using a keyboard shortcut (
CtrlP
) instead.